Taking Care of Rinn XCP and BAI instruments
Rinn XCP and BAI instruments are designed to be used in a steam
autoclave. Other forms of sterilization are not recommended.
Plastic Bite Blocks and Aiming Rings
The plastic pieces are porous and have a limited life. It is possible they
will absorb the chemicals used in sterilization and react to temperature
variation. Your plastic pieces should be replaced periodically.
- NEVER bleach your plastic pieces. Serious harm could occur to a patient or office staff from these fumes upon sterilization or use.
- Always disassemble the arm from the aiming rings and bite blocks. Because metal and plastic react differently to temperature changes, expansion and contraction of the plastic around the metal arm may cause permanent damage.
- The temperature in the autoclave should never exceed 270° F (132° C).
- Do not use ultrasonic cleaners or cold sterilizing solutions before or instead of the steam autoclave process.
Stainless Steel Aiming Arms
- Always disassemble the arms and block in a straight or parallel motion. Unusual pressure may damage the arm.
- When sterilizing, always disassemble the arm from the block and the ring. Do not bag the arm with the ring or bite-block.
|