Recommended Cleaning and Sterilization Procedure
- Disassemble instrument(s).
- Remove debris from components with hot water and soap or mild detergent.
- Load instrument(s) into autoclave and cycle per autoclave manufacturer's instructions (sterilization conditions are listed below). Do not exceed 270 F (132 C).
| Sterilizer |
Temperature/Pressure |
Time |
Steam Autoclave |
121C (250F)/15 psi (103.4 kPa) |
15 min. |
(unwrapped items) |
132C (270F)/30 psi (206.8 kPa) |
3 min. |
| |
Steam Autoclave |
121C (250F)/15 psi (103.4 kPa) |
20 min. |
(lightly wrapped items) |
132C (270F)/30 psi (206.8 kPa) |
8 min. |
| |
Steam Autoclave |
121C (250F)/15 psi (103.4 kPa) |
20 min. |
(heavily wrapped items) |
132C (270F)/30 psi (206.8 kPa) |
10 min. |
NOTE: Please carefully follow the above directions. Failure to do so may result in damage to your instruments, autoclave, or other items placed in your autoclave.