Print Close 

Recommended Cleaning and Sterilization
Procedure for Film and Sensor Holders
  1. Disassemble instrument(s).
  2. Remove debris from components with hot water and soap.
  3. Put components in sterilization pouches, and place in the middle tray of the autoclave, away from autoclave walls and heating element. Plastic parts must be in a separate pouch from metal arms, to avoid melting or warping.
  4. Cycle per autoclave manufacturer’s instructions. Do not exceed 270° F (132° C).
    – DO NOT use phenol-based glutaraldehyde.
    – DO NOT use special application or ultrasonic cleaners.
    – DO NOT CHEMICLAVE or DRY-HEAT STERILIZE.
    – DO NOT cold sterilize.

Note: Plastic parts have a limited life and should be replaced
periodically. Any method of sterilization will shorten the life of
plastic parts.

Steam Autoclave Guidelines

Temperature   Pressure Time

Unwrapped Items    
121° C (250° F) 15 psi (103.4 kPa) 15 min.
132° C (270° F) 30 psi (206.8 kPa) 3 min.

Lightly Wrapped Items
   
121° C (250° F) 15 psi (103.4 kPa) 20 min.
132° C (270° F) 30 psi (206.8 kPa) 8 min.

Heavily Wrapped Items
   
121° C (250° F) 15 psi (103.4 kPa) 20 min.
132° C (270° F) 30 psi (206.8 kPa) 10 min.


Note: Follow the above directions carefully. Failure to do so may result in damage to your instruments, autoclave, or other items placed in your autoclave.

from Safety and Infection Control in the Dental Office,
American Dental Association, Chicago, 1992; p. 44-47.