Apron Care and Maintenance
Care and Maintenance:
- Clean with soap and water as needed.
- Do not use alcohol-base cleaners.
- Store at room temperature. Do not store below 50°F (10°C) or above 100°F (37°C).
- For storage, hang aprons or lay flat. Do not fold.
- Dispose of in accordance with federal, state and local laws.
Testing
- Place an unexposed film behind the apron in the area that needs testing.
- Place the x-ray cone on the opposite side of the apron from the film and aim directly at the film.
- Expose the film at the highest level used during normal procedural exposures. Process the exposed film.
- Using a densitometer, compare the density of the exposed film to the density of a calibrated step wedge at the step equivalent to .3mm or .5mm.
- The density reading must be equal to or less than the reading at the calibrated step. If it is not, the apron is defective.
Note: a crack in the apron would appear as a dark line.